Model
|
BK-1200
|
|
Sample&Reagent System
|
Sample Positions
|
150 sample positions
|
Reagent Positions
|
180 reagent positions
|
|
Sample Volume
|
1~70μl
|
|
Reagent Volume
|
10~350μl
|
|
Probe
|
anti-collision function, liquid level detection function
|
|
Probe Washing
|
Automatic washing interior and exterior
|
|
Reagent Cooling
|
Refrigerated tray with independent switch
|
|
Reaction System
|
Temperature Control
|
Water bath 37±0.1ºC, real-time monitoring
|
Cuvettes
|
160 reusable cuvettes, optical length 6mm
|
|
Mixer Probe
|
Independent stirring
|
|
Washing
|
Automatic cuvettes washing
|
|
STAT Function
|
YES
|
|
Optical System
|
Light Source
|
12V/20W halogen lamp/fiber bragg grating optional
|
Spectrophotometry
|
Post-spectral spectrophotometry
|
|
Wavelength
|
12 wavelength from 340~800nm
|
|
Absorbance
|
0~3.5Abs
|
|
Calibration&QC
|
Calibration
|
Linear: K factor, 1-point, 2-point and multipoint linear
Non-Linear: Spline, Polygon, Logit-4P, Logit-5P |
Quality Control
|
Real-time QC, Westgard Multi Rule, Cumulative Sum Check, Twin Plot(2D), Daily QC, Monthly QC
|
|
Data Management
|
Software
|
Windows 7/8/10, 32 or 64 bit
|
LIS System
|
Bi-direction, support HL7 protocol
|
|
Interface
|
LAN port access
|
|
Printer
|
External optional, multiple reporting mode available
|
|
Working Conditions
|
Power Supply
|
AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 1500W
|
Temperature
|
15~30ºC(±2ºC)
|
|
Water Consumption
|
Deionized water: <25L/H
|
|
Humidity
|
40%~85%
|
|
Size&Weight
|
External Size (W*D*H)
|
1360*795*1155mm
|
Net Weight
|
290kg
|
|
Package Size (W*D*H)
|
1729*969*1435mm
|
|
Gross Weight
|
400kg
|