Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Biochemistry Analyzer |
Still deciding? Get samples of US$ 4600/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
200T/H Automatic Chemistry Analyzer --BK-280
Model |
BK-280 |
|
Overall Performance |
Throughput |
200Tests/Hour |
Analysis Method |
End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
|
|
|
|
Sample & Reagent Unit |
Sample Positions |
49 sample positions |
Reagent Positions |
56 reagent positions |
|
Sample Volume |
2-70μL |
|
Reagent Volume |
20-350μL |
|
Probe |
Anti-collision function,liquid level detection function |
|
Probewashing |
Automaticwashing interior and exterior |
|
ReagentCooling |
Refrigerated tray with independent switch |
|
Reaction System |
Temperature Control |
37±0.2ºC |
Cuvettes |
120 reusable cuvettes,optical length 6mm |
|
MixerProbe |
Independent stirring |
|
Washing |
Automatic cuvettes washing |
|
STAT function |
YES |
|
Optical System |
Light Source |
12V/30W halogen lamp |
Spectrophotometry |
Post-spectral spectrophotometry |
|
Wavelength |
340,405,450,480,505,546,570,600,630,700,750,800nm |
|
Absorbance |
0~3.0Abs |
|
Calibration&QC |
Calibration |
Linear: K factor, 1-point, 2-point and multipoint linear Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P,Logit-5P |
Quality Control |
Real-time QC, Westgard multi rule, Cumulative sum check, Twin Plot(2D) |
|
Data Management |
Software |
Windows 7/8/10, 32or64 bit |
LIS System |
Bi-direction, support HL7 protocol |
|
Interface |
LAN port access |
|
Printer |
External, multiple reporting mode available |
|
Working Conditions |
Power Supply |
AC220V±10%, 60/50Hz,, 110V±10%, 60Hz, 300W |
Temperature |
10~30ºC |
|
WaterConsumption |
Deionized water:5L/H |
|
Humidity |
40-85% |
|
Size & Weight |
External Size (W*D*H) |
950*603*510mm |
Net Weight |
65kg |
|
Package Size(W*D*H) |
1085*700*1012mm |
|
Gross Weight |
140kg |
Features:
1. 49 Sample positions.
2. 56 Reagent positions.
3. 120 Reaction cuvettes.
4. 200 Tests per hour.
5. Probe with anti-collision function,liquid level dectection function.