After-sales Service: | Support |
---|---|
Application: | Laboratory Apparatus |
Warranty: | 1 Year |
Detection Method: | Spectrophotometer |
Advantage: | High Sensitive |
Principle: | Classic Spectrometer |
Customization: |
---|
Suppliers with verified business licenses
Features:
* Environmental Deu-terium lamp, preventing from Ozone inhalation*Socket-Type Deu-terium lamp and Tungsten lamp
*SiO2 coating optical mirror, reduces the influence from outside* GLP self-check function:check the wavelength accuracy and
Photometry accuracy, provide test report and power-off protection*Wavelength calibration, wavelength setting, change lamp sourceand dark current calibration automatically
Basic Functions:
* Photometry
Test Abs., Transmittance and Energy by fixed wavelength.
* Quantitative
Linear fit and Linear fit through zero two modes
a. Coefficient, Standard Sample input and Standard
Sample read three modes to establish standard curve.
b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings.
c. Can save 240 group curves, can test 240 data in each curve.
d. Double wavelength, Triple wavelength test functions.
*Kinetics
Used for time course scanning or reaction rate calculations A A/t, can search all data.
* Multi-Wavelength
Can test Transmittance and Abs. with 8 different wavelengths at most.
* Scanning
User can set the scan range from 190nm to 1100nm to test the max. Abs. peak value, can do derivation, arithmetical operations to the graph.
* Biology
DNA/Protein, (Optional: UV, Lowry, BCA, CBB and Biuret)
Biobase Table Top Scanning UV/Vis Spectrophotometer
Model |
BK-S360 |
BK-S380 |
BK-S390 |
Optical System |
Single Beam, Grating 1200 lines/mm |
||
Wavelength Range |
190~1100nm |
||
Spectral Bandwidth |
1.8nm |
1nm |
0.5,1, 2,4, 5nm |
Wavelength Accuracy |
±0.3nm |
||
Wavelength Repeatability |
≤0.2nm |
||
Photometric Accuracy |
±0.002A (0~0.5Abs), ±0.004A (0.5~1.0Abs), ±0.3% T (0-100% T) |
||
Photometric Repeatability |
0.001 Abs (0 ~0.5Abs), 0.002Abs (0.5 ~1.0Abs),≤0.1% T (0-100% T) |
||
Stray Light |
≤0.04% T @ 360nm; 220nm |
||
Stability |
±0.001A/h @ 500nm |
||
Baseline Flatness |
±0.0015A |
||
Noise |
±0.0004Abs |
||
Display |
65 thousand true color 7 inch TFT LCD (480*800) |
||
Photometric Mode |
T, A, C, E |
||
Photometric Range |
0-200% T,-0.301~3.0A |
||
Detector |
Silicon Photodiode |
||
Light Source |
Deu-terium Lamp, Tungsten Lamp |
||
Input |
Membrane Keypad |
||
Output |
USB-A*2 Print and data output USB-B Connect PC |
||
Standard Accesseries |
1cm glass cuvette*4,1cm quartz cuvette*2,1cm quadruple cuvette holder, scanning software |
||
Optional Accessories |
1cm standard single slot cuvette holder, long optical path cuvette holder, cuvette, micro cuvette, test tube type absorber rack, solid sample holder, deu-terium lamp, tungsten lamp, quadruple cuvette holder, printer, four-slot water bath type constant temperature absorption pool, X-Y Adjustable Micro Absorption Pool, single bath type constant temperature absorption pool, program-controlled thermostatic absorption pool seat, 8 racks, reflection accessories |
||
Power Supply |
AC110V/220±10%, 60/50HZ |
||
Package Size(W*D*H) mm |
740*570*440 |
||
Gross Weight(kg) |
25 |
Biobase Table Top Scanning UV/Vis Spectrophotometer
Biobase Table Top Scanning UV/Vis Spectrophotometer
Suppliers with verified business licenses