Customization: | Available |
---|---|
Classification: | Biochemical Apparatus |
Type: | Hematology Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Model |
BK-3100 |
|||||
Throughput |
60 samples/hour |
|||||
Principle |
Impedance for WBC differentiation and WBC/RBC/PLT count; Colorimetric method for HGB |
|||||
Channels |
2 |
|||||
Parameters |
21 parameters(including WBC, Mid#, Lym#, Gran#, Mid%, Lym%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-SD, RDW-CV, PLT, MPV,PDW,PCT, P-LCR, P-LCC) |
|||||
Histograms |
3 histograms (including WBC, RBC, PLT histograms) |
|||||
Sample Volume |
Whole blood: 10μl, Prediluted: 20μl |
|||||
Calibration |
Manual, auto calibration |
|||||
Counting Modes |
Venous, capilary, prediluted |
|||||
Printing Model |
Auto print, manual print |
|||||
LIS |
Support Bi-directional LIS |
|||||
Language |
Chinese, English, Spanish, French, Russian, Ukrainian |
|||||
Reagent |
Diluent 20L, Hemolytic agent 500ml, Probe cleanser: 50ml |
|||||
Data Input |
9.7-inch touch screen, mouse, keyboard (optional) |
|||||
Output |
Internal printer, support external printer |
|||||
Printer paper |
57*30mm |
|||||
Interface |
4 USB ports, network port,RS232 serial port. |
|||||
Storage |
100,000 results with histogram |
|||||
Unclog |
High voltage burn, reverse high-pressure flush |
|||||
Work Temperature |
15-30°C |
|||||
Power Supply |
100~240 AC, 50/60Hz, 150VA |
|||||
Relative Humidity |
30% ~ 85 % |
|||||
Atmospheric Pressure |
70.0kPa-110.0kPa |
|||||
Carry-over rate |
Parameter |
WBC |
RBC |
HGB |
PLT |
|
CV |
≤ 1.5% |
≤ 1% |
≤ 1% |
≤ 3% |
||
Background |
Parameter |
WBC |
RBC |
HGB |
HCT |
PLT |
Background |
≤ 0.3×109/L |
≤ 0.03×1012/L |
≤ 1 g/L |
≤ 0.5% |
≤ 10×109/L |
|
Linearity |
Parameter |
Measurement range |
CV |
|||
WBC |
(0.5~10.00×109)/L |
(±0.5×109)/L |
||||
RBC |
(0.05~1.00×1012)/L (1.01~9.99×1012)/L |
(±0.05×1012)/L |
||||
HGB |
(2~70g)/L |
(±2.0g)/L |
||||
PLT |
(10~100×109)/L |
(±10.0×109)/L |
||||
Precision |
Parameter |
Whole blood (CV) |
Pre-diluted(CV) |
Measurement range |
||
WBC |
≤ 2.0% |
≤ 4.0% |
(4.0-15.0×109)/L |
|||
RBC |
≤ 1.6% |
≤ 3.0% |
(3.5-6.0×1012)/L |
|||
HGB |
≤ 1.6% |
≤ 3.0% |
(110.0-180.0g)/L |
|||
MCV |
≤1.0% |
≤2.0% |
(70.0-120.0)fL |
|||
PLT |
≤ 4.0% |
≤8.0% |
(150.0-500.0×109)/L |
|||
External Size(W*D*H) |
295*467*418mm |
|||||
Packed Size(W*D*H) |
Instrument: 404*564*616mm; Reagent: 480*480*480mm |
|||||
Net weight |
19 KG |
|||||
Gross weight |
Instrument:31KG; Reagent::25Kg |
BIOBASE Group adheres to
The Mission of "Service for Health for The Future"
BIOBASE is a new high-tech enterprise whose business covers biomedical, e-commerce, intelligent manufacturing,
scientific equipment, In Vitro Diagnosis (IVD) instrument and reagent.
With more than 5,000 employees, 22 workshops, covering a total area of 932,900 m². We have established overseas companies in USA, Dubai, India, Algeria, Singapore and other countries.
BIOBASE is specialized in medical equipment and laboratory equipment products, with more than 200 existing high-tech products.
BIOBASE Group have completed the construction of BIOBASE first industrial park in Zhangqiu, covering an area of 80,000 square meters, second industrial park in Tangye new district, covering an area of 160,000 square meters, and USA incubator in Silicon valley, covering an area of 2000 square meters.