Customization: | Available |
---|---|
After-sales Service: | Online Support |
Application: | Laboratory Apparatus |
Still deciding? Get samples of US$ 2860/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Basic Functions:
1, Photometry
Test Abs., Transmittance and Energy by fixed wavelength
2, Quantitative
Linear fit and Linear fit through zero two modes
a. Coefficient, Standard Sample input and Standard Sample read three modes to establish standard curve
b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings
c. Can save 240 group curves, can test 240 data in each curve
d. Double wavelength, Triple wavelength test functions.
3, Kinetics
Used for time course scanning or reaction rate calculations△A/t, can search all data
4, Multi-Wavelength
Can test Transmittance and Abs. with 8 different wavelengths at most
5, Scanning
User can set the scan range from 190nm to 1100nm to test the max. Abs. peak value, can do derivation, arithmetical operationsto the graph.
6, Biology
6 methods: DNA/Protein, UV, Lowry, BCA,CBB and Biuret
Model |
BK-D560 |
BK-D580 |
BK-D590 |
Optical System |
Double Beam, Grating 1200 lines/mm |
||
Wavelength Range |
190~1100nm |
||
Spectral Range |
1.8nm |
1nm |
0.5,1,2,4,5nm |
Wavelength Accuracy |
± 0.3nm |
||
Wavelength Repeatability |
≤ 0.2nm |
||
Photometric Accuracy |
± 0.002A (0~0.5Abs), ± 0.004A (0.5~1.0Abs), ± 0.3% T (0~100% T) |
||
Photometric Repeatability |
0.001Abs (0~0.5Abs), 0.002Abs (0.5~1.0Abs), ≤ 0.2% T (0~100% T) |
||
Stray Light |
≤ 0.04 T @ 360nm; 220nm |
||
Stability |
± 0.0003A / h @ 500nm |
||
Baseline Flatness |
± 0.0005A |
||
Noise |
± 0.0002Abs |
||
Display |
65 thousand true color 7 inch TFT LCD(480 *800) |
||
Photometric Mode |
T, A, C, E |
||
Photometric Range |
0~200% T,-0.301~3.0A |
||
Detector |
Silicon Photodiode |
||
Light Source |
Deuterium Lamp, Tungsten Lamp |
||
Input |
Membrane Keypad |
||
Output |
USB-A*2 Print and data output USB-B Connect PC |
||
Standard Accessories |
Scanning software,1cm standard single slot cuvette holder,1cm glass cuvette*4,1cm quartz cuvette*2 |
||
Optional Accessories |
1cm standard single slot cuvette holder, long optical path cuvette holder, cuvette, microcuvette, test tube type absorber rack, solid sample holder, deuterium lamp, tungsten lamp,quadruple cuvette holder, printer, four-slot water bath type constant temperature absorptionpool, X-Y Adjustable Micro Absorption Pool, single bath type constant temperature absonrption pool, program-controlled thermostatic absorption pool seat, 8 racks, reflectionaccessories |
||
Power Supply |
AC110V/220±10%, 60/50Hz |
||
Packing Size(mm) |
740*570*440 |
||
Gross Weight(kg) |
25 |
BIOBASE is a one stop supplier of laboratory instrument for over 20 years in Shandong Province of China. We have our own warehoue and factory to provide fast production and delivery.
1. Fast delivery, Moset models are in stock.
2. Powerful after-sale service team, our engineers can help with any after-sale problem.
3. Lots of shipment method available, air way, express, land shipment, sea shipment, we will always quote the best freight for every client.
Etc...